Abstract

Sera from 1000 women attending antenatal clinics were screened for the presence of rubella antibody by haemagglutination inhibition (HI), single radial haemolysis (SRH), and enzyme-linked immunosorbent assay (ELISA). With 925 sera concordant results were obtained with all three tests. There were 46 false-negative HI tests due to the necessity of allowing for residual non-specific inhibitors. With SRH there was one false positive and three that were negative by SRH but HI- and ELISA-positive. By ELISA 14 sera gave a false-positive result (OD 405 nm greater than or equal to 0.2). Ten of these could be eliminated by taking an OD 405 nm of 0.5 as the threshold but then another 10 sera became false negatives.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.