Abstract
A group of five subjects were fitted with truncated toric soft lenses varying in minus cylinder power from 1 to 4 D in 1 D steps at axes of 90 and 180. After 15 min of lens wear, the location of the lens was measured by three techniques: the projected slitlamp photographic technique, measurement with a protractor graticule in the slitlamp eyepiece, and alignment of the slitlamp beam with markings on the lens. Analysis of the data by a four-level analysis of variance and a t-test showed no significant difference between the three methods of measuring the toric soft lens orientation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: American journal of optometry and physiological optics
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.