Abstract

The interface pressures were measured between the sacrum, mid-lumbar spine and various support surfaces. Thirty healthy male volunteers were recruited. The spinal board, padded spinal board and vacuum stretcher were the support surfaces evaluated. We found high and potentially ischaemic pressures between the sacrum and the spinal board interface (mean 747.3 mmHg). This was reduced in the padded board (115.5 mmHg) but dramatically reduced with the vacuum stretcher (36.7 mmHg). It was also noted that no support was given to the normal lumbar lordosis by the spinal board (padded and unpadded), but support was given by the vacuum stretcher. This raises the question of how stable is an unstable spinal injury on a flat supporting surface.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.