Abstract

The advantages and limitations of the slowing down of epithermal neutrons and the thermal neutron reflection methods were compared for bulk hydrogen analysis. In both cases, the same sample dimensions were used, i.e. 10 cm diameter and 8 cm height. It was found that the sensitivity of the slowing down method is higher by a factor of ∼20 in average and the matrix effect can be neglected as compared to the neutron reflection technique. The simultaneous application of the two methods renders the determination of the H content and the C/H atomic ratio possible. The effect of spatial distribution of hydrogen in the matrix to be investigated on the number of reflected neutrons was also studied.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.