Abstract

A radiophotoluminescent glass rod dosimeter (RPL-GRD) system has recently become commercially available. The purpose of this study was to investigate the dosimetric characteristics (reproducibility, linearity, fading, energy dependence and angular dependence) of the RPL-GRD for a mailed dosimeter and to compare it with LiF-TLD powder. In this study, the model GD-301 GRD and TLD-700 were powder type used. All measurements with the exception of angular dependence were performed in a water phantom using a holder stand. The RPL-GRD has better reproducibility than the TLD for the Co-60 beam as well as for the clinical photon beam. The RPL-GRD signal was linear as a function of applied dose in the range of 0.5–3 Gy for the Co-60 gamma rays. The fading of the RPL-GRD after a received dose of 2 Gy was initially found to be within 1.7% for five months. The energy dependence of both dosimeters was found to be less than 1.6% for photon beams, but was less than 5.0% for electron beams, which was in agreement with published data. The angular dependence of the RPL-GRD was measured to be approximately 1.4% for angles ranging ±90° from the beam axis using a spherical polystyrene phantom. The measurements comparing RPL-GRD and TLD dosimetric characteristics demonstrated that the RPL-GRD is suitable for mailed dosimetry in a quality assurance (QA) audit program.