Abstract

Single-use supraglottic airway devices are now available and are intended to be comparable with the reusable LMA Classic laryngeal mask airway. We performed a randomised cross-over study comparing the Ambu AuraOnce Laryngeal Mask with the LMA Classic. Fifty patients participated in the trial. Success rates for insertion at the first attempt were similar (92% with the Ambu and 84% with the LMA Classic; p = 0.22). The volumes of air required to inflate the cuff to produce a seal were similar, but the cuff pressure was lower for the Ambu Laryngeal Mask (median (IQR [range]) 18 (10-31 [0-100] cmH(2)O) than the LMA Classic 27 (17-50 [4-90] cmH(2)O; p = 0.007). Visual analogue scores for ease of insertion were 87 (73-93 [26-97]) mm for the Ambu and 84 (60-89 [18-96]) for the LMA Classic (p = 0.017). Complications were similar in both groups. We suggest that the disposable Ambu Laryngeal Mask is an acceptable alternative to the reusable LMA Classic.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.