Abstract

This study aimed to compare the intubation effectiveness of the bébé Vie Scope™ (VieScope) and direct laryngoscopy for emergency intubation in a pediatric manikin model performed by paramedics with and without personal protective equipment for aerosol generating procedures (PPE-AGP). Participants performed endotracheal intubation using VieScope and standard Macintosh laryngoscope (MAC) in two research scenarios: (1) without PPE-AGP, and (2) with PPE-AGP. Fifty-one paramedics without any previous experience with the VieScope participated in this study. In the PPE-AGP scenario, in the VieScope group, the percentage of successful tracheal intubation on the first attempt was higher compared to the MAC group (94.1 vs. 78.4%, p = 0.031), intubation time was shorter (29.8 vs. 33.9 s, p < 0.001), and percentage of glottic opening (POGO) score was higher 91.0 vs 77.8 (p < 0.001). On the Cormack-Lehane scale, intubation with VieScope intubation was associated with higher scores rated at 1 (64.7 vs. 29.4%) than in the MAC group (p = 0.001). For intubation in the non-PPE scenario, there were no statistically significant differences between VieScope and MAC in relation to above parameters. Summarize, the bébé VieScope™ under PPE-AGP wearing conditions has proven to be a useful device for airway management in children providing better visualization of the larynx, better intubation conditions, and a higher success rate of tracheal intubation on the first attempt and reduced intubation time compared to the standard Macintosh laryngoscope.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call