Abstract

Prior to 2008, data collection from Actigraph accelerometers was only possible in the uniaxial mode. In 2009, Actigraph released the GT3X, which allows triaxial data collection. The purpose of this study was to determine if data collected by the GT3X in the uniaxial mode are comparable to its predecessor, the GT1M, under both standardized and free-living conditions. Thirty-four subjects (17 women and 17 men) provided complete data for this study. Subjects wore the accelerometers (one GT1M and one GT3X) on their waistband in line with the right and left anterior axillary lines. Each subject walked on a treadmill at speeds of 2.4, 3.2, 4.0, 4.8, 5.6 and 6.4 km h−1 for 5 min each, and then continued to wear both accelerometers for all waking hours for three consecutive days. Mean steady-state activity counts min–1 for both accelerometers were not statistically different for the standardized treadmill walking speeds and for mean minutes/day and activity counts/day for intensity classifications under the free-living condition. Based on comparisons made from both standardized walking speeds and free-living conditions, it is reasonable to compare data derived from either the GT1M or GT3X when collected in the uniaxial mode.