Abstract
The performances of three recently published leaf sequencing algorithms for step-and-shoot intensity-modulated radiation therapy delivery that eliminates tongue-and-groove underdosage are evaluated. Proofs are given to show that the algorithm of Que et al (2004 Phys. Med. Biol. 49 399–405) generates leaf sequences free of tongue-and-groove underdosage and interdigitation. However, the total beam-on times could be up to n times those of the sequences generated by the algorithms of Kamath et al (2004 Phys. Med. Biol. 49 N7–N19), which are optimal in beam-on time for unidirectional leaf movement under the same constraints, where n is the total number of involved leaf pairs. Using 19 clinical fluence matrices and 100 000 randomly generated 15 × 15 matrices, the average monitor units and number of segments of the leaf sequences generated using the algorithm of Que et al are about two to four times those generated by the algorithm of Kamath et al.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.