Abstract

Introduction  The motivation to improve accuracy and reduce complication rates in spinal surgery has driven great advancements in robotic surgical systems, with the primary difference between the newer generation and older generation models being the presence of an optical camera and multijointed arm. This study compares accuracy and complication rates of pedicle screw placement in older versus newer generation robotic systems reported in the literature. Methods  We performed a systemic review and meta-analysis describing outcomes of pedicle screw placement with robotic spine surgery. We assessed the robustness of these findings by quantifying levels of cross-study heterogeneity and publication bias. Finally, we performed meta-regression to test for associations between pedicle screw accuracy and older versus newer generation robotic spine system usage. Results  Average pedicle screw placement accuracy rates for old and new generation robotic platforms were 97 and 99%, respectively. Use of new generation robots was significantly associated with improved pedicle screw placement accuracy ( p  = 0.03). Conclusion  Accuracy of pedicle screw placement was high across all generations of robotic surgical systems. However, newer generation robots were shown to be significantly associated with accurate pedicle screw placement, showing the benefits of upgrading robotic systems with a real-time optical camera and multijointed arm.