Middle ear surgeries involve Total Ossicular Replacement Prosthesis (TORP) or Partial Ossicular Replacement Prosthesis, where all or some of the three ossicle bones are replaced. Current prostheses are dissimilar to the natural ossicles in geometry, size and only recover up to 75% sound transmission. Additive manufacturing offers complex, near-net shape geometries that allow for patient specific implants. A novel design is used to manufacture a TORP in Ti6Al4V(ELI) via Additive Manufacturing. The sound transmission level of the of the additively manufactured titanium prosthesis was measured by laser doppler vibrometry and compared to that of a Silver additively manufactured prosthesis. The sound transmission level of the additively manufactured prostheses are also compared to that of the international standards and literature using this method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.