Abstract

We compared two types of automatic non-invasive blood pressure measuring device with sphygmomanometey in 47 normotensive and 38 hypertensive women in the third trimester of pregnancy. An automatic oscillometric device (Accutor) and a volume-clamp device (Finapres) significantly underestimated the diastolic pressure as measured by the fourth Korotkoff sound using a Hawksley random zero sphygmomanometer. The mean difference between the sphygmomanometer and Accutor measurement of diastolic blood pressure was +3.1 mmHg in the normotensive women and +8.3 mmHg in the hypertensive women ( P=0.001). The mean difference between the sphygmomanometer and Finapres measurement of diastolic blood pressure was +6.1 mmHg in the normotensive women and +11.5 mmHg in hypertensive women ( P=0.003). The increased use of continuous non-invasive devices to monitor blood pressure in women with hypertension should be accompanied by sound knowledge of their limitations in this group of patients.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.