Abstract

International sensitivity index calibrations based on the W.H.O. recommended method depend on orthogonal regression analysis. As this is not readily available in statistical packages, comparison has been made with simple linear regression analysis in a study of coagulometer effects on the International Normalized Ratio (INR) at 155 European centres. Sets of seven lyophilized normal and 20 lyophilized artificially depleted abnormal plasmas were provided with five coumarin test plasmas and two European Concerted Action on Anticoagulation reference thromboplastins (low International Sensitivity Index (ISI) human and high ISI rabbit). Local ISI based on the artificially depleted lyophilized plasmas using conventional orthogonal regression gave good correction for local coagulometer effects on the human reagent and minimal correction with the rabbit reagent INR. Results were considerably worse after attempts at correction using calibration based on linear regression analysis with both reagents. The results indicate that calibration of coagulometer prothrombin time systems using simple linear regression is not appropriate.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.