Abstract
For 60Co gamma -radiation and electrons of incident energies from 3 to 35 MeV, dose measurements using ionization and FeSO4 methods are compared. If factors given by the HPA 'Practical guide to electron dosimetry' are used, results from the two methods do not differ by more than 4%. However, the ionization results based on the HPA values of CE give a mean G-value of 15.2 over this energy region. If it is assumed that photon beams producing primary electrons of a given mean energy have the same G-value as electrons of that energy, then electron G-values predicted from photon results using HPA Clambda values and the same FeSO4 equipment are about 8% higher than experimental values for 6 to 10 MeV electrons.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.