Abstract
Three image analysis software, Photoshop, Sorbfil TLC Videodensitometer software and Scion Image,were used for quantitative evaluation of TLC images of ceftriaxone sodium (CFX). Regression plots anddetection sensitivity of quantification from each TLC-image analysis were determined and compared toTLC-densitometry. TLC-image analysis method using Scion Image and Sorbfil TLC Videodensitometersoftware and TLC-densitometry showed the polynomial regression data with good relationship (R2 > 0.99)over the concentration range of 1-8 mg/spot whereas the analysis of TLC images with Photoshop showedgood polynomial regression plots (R2 > 0.99) at higher concentration range, 4-10 mg/spot. For detectionsensitivity, LOD and LOQ determined from TLC-image analysis using Sorbfil TLC Videodensitometer werecomparable to the values from TLC-densitometry. Scion Image and Photoshop was found to be less sensitive.The use of Sorbfil TLC Videodensitometer software for TLC image analysis could be further applied forrapid determination of CFX in bulk drug and dosage forms. Key Words: TLC; Image analysis software; Ceftriaxone sodium
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Silpakorn University Science and Technology Journal
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.