Abstract

PurposeTo evaluate factors associated with differences in intraocular pressure (IOP) readings between iCare and Goldmann applanation tonometry (GAT) in established glaucoma patients.MethodsThis retrospective comparative study included clinical data of 350 eyes from 350 established glaucoma patients who had iCare and GAT IOP measured by an ophthalmic technician and a glaucoma specialist, respectively. The main outcome measure was the difference in IOP measurements of the right eyes with iCare and GAT.ResultsThe intraclass correlation coefficient (ICC) between GAT and iCare was 0.90. The mean IOP difference between tonometers was − 0.18 ± 2.89 mmHg. Bland–Altman plots indicated a 95% limit of agreement of − 5.8 to 5.5 mmHg. Central corneal thickness (CCT) and age were significantly correlated with the difference in IOPs of the iCare and GAT. GAT-IOP and age were significantly associated with the absolute difference in measured IOP of the two tonometers. The difference in measurements was not significantly associated with prior glaucoma surgery, average global index of optical coherence tomography, axial length, technician years of experience and certification, and IOP range.ConclusionAlthough there is good agreement between the iCare and GAT mean values, these devices are not interchangeable in glaucoma patients due to the wide range of the limit of agreement.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.