Flow D- and E-speed group dental X-ray films (DX-58 and EX-58) have recently been introduced to the market. By means of perceptibility curves these films were compared with commonly used dental X-ray films (Kodak Ultraspeed and Ektaspeed). No major differences between the films were found with regard to contrast and the subjectively assessed number of small contrast differences. The exposure for EX-58 had to be lowered by 66% and Ektaspeed by 39% compared with Ultraspeed film to obtain the same density. DX-58 and Ultraspeed were of the same sensitivity.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.