Abstract

AIM: To compare the performance of a direct digital mammography system with normal-view and magnified-view conventional screen-film methods using quality control phantoms.MATERIALS AND METHODS: Using a Siemens Mammomat®3000 and an Opdima®digital spot imaging and biopsy attachment, film and direct digital images of two phantoms [DuPont and TOR (MAM)] were obtained under normal operating conditions. These were assessed by three groups of observers with differing expertise — radiologists, radiographers and medical physicists. Each observer was asked to compare the direct digital image with films taken in standard view and magnified view, providing scores for object visibility and confidence. For the digital images, observers were allowed to vary the image presentation parameters.RESULTS: Both phantoms showed that overall the direct digital view and the magnified view film performed significantly better (P<0.05) than standard view film. For certain small or low contrast objects the differences became very highly significant (P<0.001).CONCLUSION: Only the TOR (MAM) phantom showed any significant difference between digital and magnified modalities, with magnified views performing better for fine, faint filaments and digital acquisition better for low contrast objects. Almost no difference existed between the three observer groups.Undrill, P. E. (2000).Clinical Radiology53 , 782–790.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.