Abstract

Non-invasive beat-to-beat mean arterial pressure (MAP) in finger arteries recorded by the differential oscillometric device was compared with MAP recorded invasively from A. radialis in 22 patients after cardiac surgery. Based on all 132 paired measurements, the MAP values measured at the radial artery were 2.7 ± 4.9 mmHg higher than those measured on fingers. Among 22 patients there were 8 patients receiving inotropic support, their difference being 2.1 ± 5.6 mmHg. The present study revealed that the mean discrepancy between the invasive radial pressure and finger pressure was small; however, patient data sets showed marked variability in average pressure differences when examined individually.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.