Abstract

A quantitative respirator fit test system based on controlled negative pressure was evaluated by comparison testing with a computerized aerosol fit test system. Experiments ranged from multiple sequential tests of a single subject wearing a respirator equipped with a series of fixed leaks to sequential fit tests of 125 U.S. Air Force personnel using both systems. Throughout each test phase, measured negative pressure fit factors were consistently more conservative and less variable than aerosol fit factors. Comparison of subject and fixed leak fit factors indicated significant loss of aerosol during subject fit tests. Negative pressure system results did not show any effect from subject-related losses.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.