Abstract

This clinical study compared induction time, consumed anaesthetic dose, and haemodynamic and recovery profiles when using a new type of multifunctional anaesthesia machine (Zeus) in semi-closed or closed circuit modes. Sixty female patients undergoing gynaecological surgery were randomly assigned to three groups and received desflurane anaesthesia through a semi-closed circuit (SCC) at fresh gas flow rates of 4 l/min (SCC 4 l/min) or 2 l/min (SCC 2 l/min), or through a closed circuit (CC). Anaesthesia was maintained at the minimum alveolar concentration for blocking the adrenergic response to painful stimulus (MAC(BAR)) (4.6% end-tidal desflurane) during each operation. The time required to reach MAC(BAR) was significantly shorter and the dose of desflurane was significantly smaller in the CC group compared with the other groups. There were no differences in haemodynamic and recovery profiles between the groups. It is concluded that the CC mode allowed a faster and more reliable induction, lower anaesthetic consumption and stable haemodynamic and recovery profiles.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.