Abstract

Cochlear implant manufacturers utilize different artifact-reduction methods to measure electrically evoked compound action potentials (ECAPs) in the clinical software. Two commercially available artifact-reduction techniques include forward masking (FwdMsk) and alternating polarity (AltPol). AltPol assumes that responses to the opposing polarities are equal, which is likely problematic. On the other hand, FwdMsk can yield inaccurate waveforms if the masker does not effectively render all neurons into a refractory state. The goal of this study was to compare ECAP thresholds, amplitudes, and slopes of the amplitude growth functions (AGFs) using FwdMsk and AltPol to determine whether the two methods yield similar results. ECAP AGFs were obtained from three electrode regions (basal, middle, and apical) across 24 ears in 20 Cochlear Ltd. recipients using both FwdMsk and AltPol methods. AltPol waveforms could not be resolved for recipients of devices with the older-generation chip (CI24R(CS); N = 6). Results comparing FwdMsk and AltPol in the CI24RE- and CI512-generation devices showed significant differences in threshold, AGF slope, and amplitude between methods. FwdMsk resulted in lower visual-detection thresholds (p < 0.001), shallower slopes (p = 0.004), and larger amplitudes (p = 0.03) compared with AltPol. Results from this study are consistent with recent findings showing differences in ECAP amplitude and latency between polarities for human CI recipients. When averaged, these differences likely result in a reduced ECAP response with AltPol. The next step will be to separate the effects of artifact-reduction method and stimulus polarity to determine the relative effects of each.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call