Abstract

Aim : To assess the accommodative amplitudes of three different foldable acrylic intraocular lenses (IOLs) by measuring with photorefractor Plusoptix CR03 (Plusoptix GmbH, Nuernberg, Germany). Methods: Fifty five pseudophakic eyes of 46 patients (mean age 65.25±10.63 years) were analyzed one month after uneventful phacoemulsification and IOL implantation surgery. Rayner Centerflex (C-Flex) (n=17), AcyrSof MA60BM (n=17), Sensar AR40e (n=21) IOLs were implanted. An objective, dynamic measurement technique with R mode of Plusoptix CR03 device was used to measure accommodation. Accommodation was measured at near distance (0.33m) and far distance (5m). Results: Accommodation amplitudes (mean ± SD) measured with photorefractor at near distance in Rayner, AcyrSof and Sensar groups were 1.45±0.76 diopter (D), 2.82±1.88 D, 2.44±1.02 D, respectively. Accommodative response of AcyrSof and Sensar IOLs was greater than Rayner IOL significantly (p=0.027 and p=0.004, respectively). At far distance,

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.