Abstract

Loading in total knee arthroplasty (TKA) is multifactorial and dependent on alignment, ligament balance, patient, and implant factors. Abnormal loading has been linked to clinical failure; however, the respective contribution of each factor to failure is not well known. This study defined the effect of metal backing on loading patterns in the proximal tibia. Composite tibiae were implanted with metal-backed and all-polyethylene Anatomic Graduated Component TKA tibial components (Biomet, Inc, Warsaw, Ind) and coated with photoelastic material allowing full-field dynamic strain quantification. In simulated varus loading distributions, significant increases in measured strain were observed ranging from 40% to 587% for all-polyethylene vs metal-backed tibial components. Higher observed strains in the proximal tibia observed with all-polyethylene tibial components could possibly explain increased clinical failure rates observed with this TKA design.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.