Abstract

IntroductionTotal 25-hydroxyvitamin D [25(OH)D] is the most reliable indicator of vitamin D status. In this study, we compared two automated immunoassay methods, the Abbott Architect 25-OH Vitamin D assay and the Roche Cobas Vitamin D total assay, with the liquid chromatography-tandem mass spectrometry (LC-MS/MS).Materials and methodsOne hundred venous blood samples were randomly selected from routine vitamin D tests. Two of the serum aliquots were analyzed at the Abbott Architect i2000 and the Roche Cobas 6000’s module e601 in our laboratory within the same day. The other serum aliquots were analyzed at the LC-MS/MS in different laboratory. Passing-Bablok regression analysis and Bland-Altman plot were used to compare methods. Inter-rater agreement was analyzed using kappa (κ) analysis.ResultsThe Roche assay showed acceptable agreement with the LC-MS/MS based on Passing-Bablok analysis (intercept: -5.23 nmol/L, 95% CI: -8.73 to 0.19; slope: 0.97, 95% CI: 0.77 to 1.15). The Abbott assay showed proportional (slope: 0.77, 95% CI: 0.67 to 0.85) and constant differences (intercept: 17.08 nmol/L; 95% CI: 12.98 to 21.39). A mean bias of 15.1% was observed for the Abbott and a mean bias of -14.1% was observed for the Roche based on the Bland-Altman plots. We found strong to nearly perfect agreement in vitamin D status between the immunoassays and LC-MS/MS. (κ: 0.83 for Abbott, κ: 0.93 for Roche) using kappa analysis.ConclusionBoth immunoassays demonstrated acceptable performance, but the Roche Cobas assay demonstrated better performance than the Abbott Architect in the studied samples.

Highlights

  • Total 25-hydroxyvitamin D [25(OH)D] is the most reliable indicator of vitamin D status

  • The Roche assay showed acceptable agreement with the LC-MS/MS based on Passing-Bablok analysis

  • We found strong to nearly perfect agreement in vitamin D status between the immunoassays and LC-MS/ MS. (κ: 0.83 for Abbott, κ: 0.93 for Roche) using kappa analysis

Read more

Summary

Materials and methods

One hundred venous blood samples were randomly selected from routine vitamin D tests. Two of the serum aliquots were analyzed at the Abbott Architect i2000 and the Roche Cobas 6000’s module e601 in our laboratory within the same day. The other serum aliquots were analyzed at the LC-MS/MS in different laboratory. Passing-Bablok regression analysis and Bland-Altman plot were used to compare methods. Inter-rater agreement was analyzed using kappa (κ) analysis

Results
Conclusion
Study design
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.