Abstract

This paper presents the results of a comparison of active/reactive power meter calibrations between the National Institute of Standards and Technology and National Research Council. The comparison was implemented using a transfer standard consisting of a highly stable commercial sampling-type power/energy meter. Active and reactive power measurements were made at 120 V, 5 A, 50 Hz, and 60 Hz. For active power, the measurements were made at applied current phase angles of 0°, +60°, and -60°. For reactive power, the measurements were made at applied current phase angles of +60°, +90°, -60°, and -90° . The results of the comparison indicate agreement to within the stated uncertainties of the participants.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.