Abstract

Sternum closure after open heart surgery is typically done with steel wires. Final approximation of sternal parts and connection is achieved by twisting the ends of the wire and bending the twisted assembly towards the sternum in order to minimize outward protrusion. Though this routine procedure is highly effective, some failures do occur, e.g. due to wire fracture. Fatigue fracture of the wires, e.g. due to coughing implies a failure risk. An alternative development is to make cables from gel spun Ultra High Molecular Weight Poly Ethylene (UHMWPE) fibres, such fibres are extremely strong, yet flexible, and if made as a very pure grade, they are highly bio compatible. The optimal connection technique will be different from that of steel. Connection will rather be with knotting than twisting. A new sternum closure and fixation technique has been developed for the sternum. Additionally, a testing technique was developed, for a connection of simulated sternum parts, using different materials according to their respective optimal connection method and subsequently testing the mechanical properties of the connection. Substantial differences were observed. The mechanical behaviour of twisted steel wire connection showed more scatter than the knotted UHMWPE cables and some initial slack was sometimes present in the twisted cables. The maximum attainable force in the steel wires was determined by “untwisting” due to the external load. The maximum force in the UHMWPE cables was determined by the knot strength, either slipping for small knots, or breaking of the cables at the knots for slip-improved knots. The maximum force on the knotted UHMWPE cables was substantially larger than the maximum force on the twisted steel wires. Fatigue tests were performed on both the steel solution and the UHMWPE cables solution. The performance was about similar, although the simulated sternum opening was smaller for the UHMWPE cables at higher load levels. Summarizing, the UHMWPE cables show two advantages namely higher maximum load and more reproducible mechanical behaviour due to less scatter in the mechanical behaviour. On the other hand, the connection by knotting UHMWPE cables is somewhat more elaborate than the simple twisting connection of steel wires.

Highlights

  • Open heart surgery may be inevitable if other treatments of heart disease, like angioplasty, fail

  • The results showed some scatter, so two typical load displacement diagrams are presented per material and load level

  • It is observed that the twisted steel wire connections sometimes show some “slack”

Read more

Summary

Introduction

In-vivo broken wire that was removed from the patient, about five month after surgery. The flexural stiffness and hardness of the steel wire will cause inhomogeneous load distribution on the sternum, after closing and tightening. A possible solution for these problems may be to close the sternum with polymer cables. Such cables may be less sensitive to fatigue and they are extremely flexible, a very homogeneous load distribution on the sternum may be expected. Considering the occurring loads, it is clear that the tensile strength of the polymer cable must about equalize the strength of the steel wire. Cables made from gel-spun UHMWPE do offer that high tensile strength. Dyneema Purity® is a specially produced grade with extreme purity, making it suitable as an implant material.

Materials and Experiments
Findings
Discussion and Conclusions

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.