Abstract

Patent hemostasis (PH) is essential for preventing radial artery occlusion (RAO) after trans-radial procedures; however, it remains unclear how it should be obtained. The aim of this multicenter randomized study was to evaluate whether the use of an adjustable device (AD), inflated with a pre-determined amount of air (AoA), was more effective than a non-adjustable device (non-AD) for achieving PH, thereby reducing the incidence of RAO. We enrolled a total of 480 patients undergoing transradial procedure at 3 Italian institutions. Before the procedure, a modified Reverse Barbeau Test (mRBT) was performed in all patients to evaluate the AoA to be eventually inflated in the AD. After the procedure, patients were randomized into 2 groups: (1) AD Group, using TR-Band (Terumo) inflated with the pre-determined AoA; and 2) non-AD Group, using RadiStop (Abbott). An RBT was performed during compression to demonstrate the achievement of PH, as well as 24 hours later to evaluate the occurrence of RAO. PH was more often obtained in the AD Group compared with the non-AD Group (90% vs 64%, respectively, P less than .001), with no difference in terms of bleedings. RAO occurred more often in the non-AD Group compared with the AD Group (10% vs 3%, respectively, P less than .001). Of note, mRBT was effective at guiding AD inflation and identifying high-risk patients in whom PH was more difficult to obtain. The use of AD, filled with a predetermined AoA, allowed PH significantly more often compared with non-AD, providing a significantly reduced incidence of RAO.