Abstract

This cross-sectional pilot-study investigated the reproducibility of the LDI (Moor-LDI-B2; Moor Instruments) and the chromameter (Minolta chromameter CR-300) when used in scar assessment. Twenty-seven scars in 14 subjects were included between January and June 2003. One observer performed two times both measurements with 10 min apart. The intra-observer agreement is quantified by means of the intra-class correlations (ICC) and the standard errors of measurement (SEM) for both the LDI and the chromameter. Ignoring one outlier, the ICC of the LDI = 0.856 and the SEM = 34.56. The chromameter shows a better reproducibility with an ICC of 0.93 and a SEM of 0.79. This pilot-study with a limited number of measurements shows a moderate reproducibility of the LDI compared to the chromameter measurements, in the assessment of respectively flux and redness in scars.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.