Abstract
This cross-sectional pilot-study investigated the reproducibility of the LDI (Moor-LDI-B2; Moor Instruments) and the chromameter (Minolta chromameter CR-300) when used in scar assessment. Twenty-seven scars in 14 subjects were included between January and June 2003. One observer performed two times both measurements with 10 min apart. The intra-observer agreement is quantified by means of the intra-class correlations (ICC) and the standard errors of measurement (SEM) for both the LDI and the chromameter. Ignoring one outlier, the ICC of the LDI = 0.856 and the SEM = 34.56. The chromameter shows a better reproducibility with an ICC of 0.93 and a SEM of 0.79. This pilot-study with a limited number of measurements shows a moderate reproducibility of the LDI compared to the chromameter measurements, in the assessment of respectively flux and redness in scars.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.