Abstract

A numerical approach was proposed to investigate the wear behavior occurred in the artificial hip joints in this paper. In the numerical simulations, the wear coefficients taken from pin-on-disk tests were introduced into the wear analysis model to assess the wear rates of polyethylene acetabular cups against metallic or ceramic femoral heads. For the established material combinations, different values of polyethylene wear rates were obtained respectively, which were not necessarily the realistic one as expected in vivo but could be confirmed after further discussion on the wear mechanism involved in wear tests. Current results indicated that the polyethylene/ceramic couples represented better wear performances than the polyethylene/metal couples. Furthermore, the ratio of wear rates for polyethylene cups against alumina and the metallic femoral heads was 0.5, which agreed well with that deduced from clinical studies or laboratory hip simulators. It is obvious that these comparable wear behaviors observed from clinics or laboratory studies also can be found by means of the numerical simulation.