Abstract

Ganglioside (G(M1)) micelles have been studied by means of three different techniques: fluorescence correlation spectroscopy (FCS), electronic energy transfer, as monitored by time-resolved fluorescence spectroscopy, as well as static and dynamic light scattering. The aggregation numbers obtained, 168 +/- 4, remain constant over a wide range of G(M1) concentrations (0.764-156 muM), are very consistent when using different donor-acceptor energy transfer pairs and have served as reference values in tests of the FCS method. It is recommended to calibrate the focal volume by using known dye concentrations. For this the rhodamine dye, 5-TAMRA, turns out to be most suitable. It is also shown that FCS provides correct values of the aggregation numbers, provided that the focal volume is calibrated by using updated values of the diffusion constant of Rhodamine 6G. These results also support recent methodological advances in FCS.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.