Abstract

The aim of this study was to assess the primary stability of dental implants by resonance frequency analysis (RFA) using the Osstell™ and Osstell Mentor™ devices and to investigate the reproducibility and comparability of the results obtained with these devices. Twenty-four Straumann implants (Straumann AG, Basel, Switzerland) were placed in the anterior mandible of 12 fresh edentulous human cadaver mandibles. The implant stability quotients (ISQs) were measured with the Osstell™ and Osstell Mentor™ when implants were inserted at 50% of their length and following their complete insertion. The Osstell™ device measured lower scores compared with the Osstell Mentor™. This was significant for the full position (mean difference = 9·9), t (11) = 7·4, P < 0·001 and for the halfway position (mean difference = 5·9), t (11) = 2·41, P = 0·03. In conclusion, the Osstell™ produced relatively lower ISQ scores than the Osstell Mentor™.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.