Abstract

ObjectivesThe aim of this study was to determine the diagnostic accuracy of cone beam computed tomography (CBCT) in the detection of approximal and occlusal noncavitated carious lesions.MethodsA total of 135 extracted human posterior teeth were used. They were radiographed using the following: conventional film (Kodak Insight), a digital system (PSP plates Digora Optime, Soredex), and a CBCT system (NewTom 3G, Quantitative Radiology). All the images were assessed by two independent observers twice. Receiver operating characteristic analysis (Az) was used.ResultsNewTom 3G 9″ CBCT was statistically inferior to conventional film and a digital system for the detection of approximal caries. NewTom 3G 9″ had significantly higher Az values than PSP plate (p < 0.05), but there were no statistically significant differences between the Az values for CBCT and film (p > 0.33) for occlusal surface caries.ConclusionThe diagnostic accuracy of all three tested radiographic systems is low, and it is necessary to use other methods to improve early caries detection.Clinical relevanceCBCT has found a wide application in different fields of dentistry. The report from a CBCT examination performed for any of these reasons should include all abnormal findings, and the detection of noncavitated caries lesions is especially important because it facilitates the use of noninvasive treatment.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call