Abstract

Background and objectiveExtracting the mechanical behaviors of bioprosthetic aortic valve leaflets is necessary for the appropriate design and manufacture of the prosthetic valves. The goal of this study was to opt a proper tissue for the valve leaflets by comparing the mechanical properties of the equine, porcine, and donkey pericardia with those of the bovine pericardium and human aortic valve leaflets. MethodsAfter tissue fixation in glutaraldehyde, the mechanical behaviors of the pericardial tissues were experimentally evaluated through computational methods. The relaxation tests were performed along the tissue fiber direction. The Mooney-Rivlin model was utilized to describe the hyperelastic behavior of the tissues at the ramp portion. The viscous behaviors at the hold portion were extracted using the Fung quasi-linear viscoelastic (QLV) model. Furthermore, the extracted parameters were used in the modeling of the bovine, equine, porcine, and donkey pericardia through finite element analysis (FEA). ResultsBased on the results, relaxation percentages of the equine, donkey, and bovine pericardia were greater than that of the porcine pericardium and similar to the native human aortic valve leaflets. Indeed, the equine and donkey pericardia were found more viscous and less elastic than the porcine pericardium. Compared with the porcine pericardium, the mechanical properties of the equine and donkey pericardia were rather closer to those of the native human leaflets and bovine pericardium. The computational analysis demonstrated that the donkey pericardium is preferable over other types of pericardium due to the low stress on the leaflets during the systolic and diastolic phases and the large geometric orifice area (GOA). ConclusionThe donkey pericardium might be a good candidate valve leaflet material for bioprosthetic aortic valves.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call