Abstract

We present here a comparative study of atomic force microscope (AFM) imaging in contact mode when either the cantilever carrying the probing tip or the sample is excited at ultrasonic frequencies. The cantilever or the sample can be excited by piezoelectric transducers attached to them. When the AFM tip is in contact with the sample surface the contact resonance curve depends on the local tip–sample contact stiffness. By measuring the contact resonance as a function of position one can image the local stiffness of the sample surface. We will report here imaging carried out on piezoelectric material and thin metal film using both the modes. The comparison shows that both give similar results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.