Abstract

In this study, the calculations and results of acoustic voice analysis as calculated by two different analysis systems (Doctor Speech (DRS), Tiger Electronics, Neu-Anspach, Germany, and Computerized Speech Lab (CSL), Kay Elemetrics Corporation, Lincoln Park, NJ) are compared. A group of 120 normal voices was selected for analysis of the objective parameters: fundamental frequency (F(0)), variation of F(0) (F(0)SD), jitter, shimmer, and harmonics-to-noise ratio (HNR). The subject group was a random selection of normal voices of adults. The aim of this comparison was to find determined differences and similarities in data measurements between both systems to make data transfer possible. A significant correlation was found for F(0), HNR, and shimmer relative. The correlation for jitter (relative and absolute) and F(0)SD was weak. DRS and CSL are not comparable in absolute figures, but their judgment against normative data is identical. Further research is necessary to explore the affect on pathological voices or child voices.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.