Abstract

Despite the benefits of endoscopic nasobiliary drainage (NBD) in endoscopic retrograde cholangiopancreatography (ERCP), post-ERCP pancreatitis (PEP) and nose/throat discomfort can result. We aimed to determine whether the use of a smaller catheter alleviates these complications. A randomized, controlled trial at a tertiary care center compared 4 Fr and 6 Fr NBD catheters; 165 ERCP patients with naïve papillae were randomly assigned to a catheter-size group. The prevalence of PEP was significantly lower in the 4 Fr group (3.7%; 3/82) than in the 6 Fr group (15.7%; 13/83; P = 0.019). No spontaneous catheter displacement occurred within 24 h. Discomfort visual analog scores were 2.6 and 4.3 in the 4 Fr and 6 Fr groups, respectively (P = 0.0048) on procedure day; on the following day, the scores were 2.3 and 3.6 (P = 0.028). Bile output was 16.3 mL/h and 21.4 mL/h in the 4 Fr and 6 Fr groups (P = 0.051). On obstructive jaundice subgroup analysis, bile drainage was 19.2 mL/h and 22.1 mL/h in the 4 Fr and 6 Fr groups (P = 0.40). The 4 Fr group required 5.6 days to reduce bilirubin levels versus 6.1 days in the 6 Fr group (P = 0.51). In patients with naïve papillae, lower rates of PEP and less nose/throat discomfort are associated with the use of 4 Fr NBD catheters. In patients with obstructive jaundice, 4 Fr and 6 Fr catheters are comparable with regard to bile output and bilirubin level reduction.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call