Atomic absorption technique is used to determine losartan potassiumthrough the quantitative detection of potassium (K) element in the drug.the extraction was carried out by acid digestion and the percentagerecovery is 97.2%, the error percentage is -2.8, detection limit is 0.038 μg/ml, and the relative standard deviation did not exceed 2.58%. The resultswere compared with UV-Visible technique method with percentagerecovery of 96.5%, detection limit is 0.93 μg / ml and the relativestandard deviation was not more than 1.55%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.