Abstract
Introduction Blood glucose estimation is required to be done rapidly or even on hourly basis for effective dosing of insulin in diabetic patients admitted in intensive care units (ICUs). Glucose test strip precision is usually considered clinically acceptable if the difference from the reference value (lab method) does not exceed 15%. To use glucometers in ICU settings the correlation between capillary blood glucose and venous blood glucose is essential. Materials and Methods This study was a cross sectional, descriptive study which was carried out in a large tertiary care teaching hospital in Pune, from 01 Nov 2015 to 31 Oct 2016. After excluding all the confounding factors, 65 patients were included in this study. The laboratory results and glucose strip results were tabulated in a master chart as sample 1(capillary/venous- time of admission), sample 2(capillary /venous – 24 hrs), sample 3 (capillary /venous – 48 hrs) and also segregated into less than 250 mg/dl and more than 250 mg/dl for determining the correlation and agreement between the capillary and venous blood glucose levels. Results A good correlation existed between the CBG and VBG of sample 1 (R2 =0.995, p
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.