Abstract

The pressure fields of two different high intensity focused ultrasound (HIFU) transducers operated in burst mode were measured at acoustical power levels of 25 and 50 W (continuous wave equivalent) with three different hydrophones: A fiber-optic displacement sensor, a commercial HIFU needle hydrophone, and a prototype of a membrane hydrophone with a protective coating against cavitation effects. Additionally, the fields were modeled using a freely available simulations software package. The measured waveforms, the peak pressure profiles, as well as the spatial-peak temporal-average intensities from the different devices and from the modeling are compared and possible reasons for differences are discussed. The results clearly show that reliable pressure measurements in HIFU fields remain a difficult task concerning both the reliability of the measured values and the robustness of the sensors used: Only the fiber-optic hydrophone survived all four exposure regimes and the measured spatial-peak temporal-average intensities varied by a factor of up to 1.5 between the measurements and the modeling and between the measurements among themselves.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.