Abstract

Aim: This in vivo study was done to correlate the values of three different electronic apex locators: Root ZX, iPex, and Apex ID, with the radiographic working length (WL). Materials and Methods: Nineteen maxillary incisors with completely formed root apices indicated for endodontic treatment were chosen. After access preparation, coronal flaring and pulp extirpation, electronic measurement of WL was done with a No: 15 K file using the three devices on each tooth. A WL radiograph was then taken with the file placed within the canal and the length determined by following Ingle's method. The data were then analyzed using the correlation coefficient and Z -test. Results: Root ZX showed the maximum correlation ( r = 0.9881) with radiographic WL followed by Apex ID ( R = 0.9731) and iPex ( R = 0.9508). Root ZX had a statistically significant higher correlation with radiographic WL in comparison with iPex ( P Conclusion: Maximum correlation of readings with the radiographic length was shown by Root ZX followed by Apex ID and then by iPex.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.