Abstract

Tritium (3H) in natural waters is a powerful tracer of hydrological processes, but its low concentrations require electrolytic enrichment before precise measurements can be made with a liquid scintillation counter. Here, we describe a newly developed, compact tritium enrichment unit which can be used to enrich up to 2L of a water sample. This allows a high enrichment factor (>100) for measuring low 3H contents of <0.05TU. The TEU uses a small cell (250mL) with automated re-filling and a CO2 bubbling technique to neutralize the high alkalinity of enriched samples. The enriched residual sample is retrieved from the cell under vacuum by cryogenic distillation at −20°C and the tritium enrichment factor for each sample is accurately determined by measuring pre- and post- enrichment 2H concentrations with laser spectrometry.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.