Abstract

Synchrotron Radiotherapy (SyncRT) is a preclinical radiation treatment which delivers synchrotron x-rays to cancer targets. SyncRT allows for novel treatments such as Microbeam Radiotherapy, which has been shown to have exceptional healthy tissue sparing capabilities while maintaining good tumour control. Veterinary trials in SyncRT are anticipated to take place in the near future at the Australian Synchrotron's Imaging and Medical Beamline (IMBL). However, before veterinary trials can commence, a computerised treatment planning system (TPS) is required, which can quickly and accurately calculate the synchrotron x-ray dose through patient CT images. Furthermore, SyncRT TPS's must be familiar and intuitive to radiotherapy planners in order to alleviate necessary training and reduce user error. We have paired an accurate and fast Monte Carlo (MC) based SyncRT dose calculation algorithm with EclipseTM, the most widely implemented commercial TPS in the clinic. Using EclipseTM, we have performed preliminary SyncRT trials on dog cadavers at the IMBL, and verified calculated doses against dosimetric measurement to within 5% for heterogeneous tissue-equivalent phantoms. We have also performed a validation of the TPS against a full MC simulation for constructed heterogeneous phantoms in EclipseTM, and showed good agreement for a range of water-like tissues to within 5%–8%. Our custom EclipseTM TPS for SyncRT is ready to perform live veterinary trials at the IMBL.

Full Text
Published version (Free)