Abstract

Twenty-four laboratories participated in a collaborative study to calibrate a replacement for the 4th International Standard for Unfractionated Heparin (82/502). Both candidate materials A and B, gave excellent intra- and inter-laboratory variations (majority of mean %gcv <10%) when assayed against the 4th International Standard. No major differences of potency estimates were found between methods, although the USP method generally gave lower potencies than the other methods and candidate B gave a greater variation between methods than A. Overall, this study showed that the differences between the candidates are marginal. Based on its narrower molecular weight profile, higher specific activity and slightly lower inter-method variation, candidate A, 97/578, was proposed and accepted in October, 1998, by the Expert Committee on Biological Standardisation of the World Health Organisation to be the 5th International Standard for Unfractionated Heparin with an assigned potency of 2031 IU/ampoule.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.