Abstract
A system for the standardization of alpha-gamma or electron-X radionuclide emitters is described. The system consists of one or two surface barrier detectors for alpha or electron detection which are coupled to thin-window NaI(Tl) crystals suitable for low-energy X- or gamma-ray detection. The performance of the system has been verified by the standardization of 241Am, 137Cs and 109Cd solutions. The activity has been obtained using the extrapolation method applied to the 4π α-γ and 2π e-X coincidence techniques. The surface barrier detection efficiency was varied by placing absorbers over the radioactive source or by changing the source-to-detector distance. The results were compared to those obtained using conventional absolute systems based on gas-flow and pressurized 4π proportional counters, or using radioactive solutions standardized in international comparisons sponsored by the Bureau International des Poids et Mesures, France. The expected and measured activities agreed within the experimental uncertainties, which were: 0.2% for 241Am, 0.7% for 137Cs and 0.6% for 109Cd.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.