Abstract

This paper describes a new method to determine the phase angle errors of ac shunts by measuring the inductance and distributed capacitance. A 1-Ω shunt of coaxial design has been developed as the time constant standard. A coaxial inductor with identical structure as the time constant standard has been developed. A four-terminal mutual inductor has also been built for the measurement of the inductance of the time constant standard. The method is based on the use of a binary inductive current divider to compare the inductor with the mutual inductor. The standard uncertainty of the inductance measurement of the time constant standard is within 28 pH at frequencies up to 200 kHz. The expanded uncertainty of the phase angle errors is from 0.4 μrad at 1 kHz to 80.0 μrad at 200 kHz at 1 A.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.