Abstract
The worldwide prevalence of epilepsy is approximately 1%, and 25% of epilepsy patients cannot be treated sufficiently by available therapies. Brain stimulation with closed-loop seizure control has recently been proposed as an innovative and effective alternative. In this paper, a portable closed-loop brain computer interface for seizure control was developed and shown with several aspects of advantages, including high seizure detection rate (92-99% during wake-sleep states), low false detection rate (1.2-2.5%), and small size. The seizure detection and electrical stimulation latency was not greater than 0.6 s after seizure onset. A wireless communication feature also provided flexibility for subjects freeing from the hassle of wires. Experimental data from freely moving rats supported the functional possibility of a real-time closed-loop seizure controller.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.