Abstract

The strain-gauge Codman MicroSensor intracranial pressure (ICP) transducer has shown consistently good laboratory performance. To assess the practical performance of the system in patients following acute brain injury, 10 patients were fitted with a MicroSensor and a second ICP monitor. In five cases this was a fibre-optic transducer and in five cases an intraventricular fluid-filled device. Paired ICP values were recorded every 5 min. ICP values ranged from 0 to 31 mmHg. Altman-Bland plots showed that individual readings could differ by as much as 9 mmHg. Further analysis showed that much of this disagreement could be explained by a constant offset on each occasion. Comparison traces of ICP in individual patients show high agreement in timing and size of changes. The unexplained constant offset leads to uncertainty about the true ICP. Treatment decisions are often based upon absolute levels of ICP and patient care may therefore differ depending upon the monitor used.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.