Abstract
This paper discusses shortcomings of conventional Holter monitoring in paced patients and describes a new technique which permits reliable detection of intermittent pacemaker malfunction and counts pacemaker activity during the recording period. Evaluation of the system of 64 consecutive patients revealed 15 with unsuspected episodic pacemaker dysfunction.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.