Abstract
ABSTRACTBackground: The re-usable Classic laryngeal mask airway (LMA®) is widely used. There are concerns regarding the transmission of pathogens. Disposable airway devices provide a cost-effective alternative. We performed a side by side clinical comparison of these devices applicable to the South African context.Methods: Adult ASA 1–3 patients (30–100 kg) presenting for elective peripheral surgery in Tygerberg Academic Hospital were randomised to receive the gold standard Classic LMA®, or one of four disposable devices. They all received a standardised anaesthetic with propofol, fentanyl and isoflurane in 40% O2/N2O. Insertion technique, mask sizes and maximum cuff volumes were per manufacturer's instructions. The cuff was inflated to achieve an adequate airway seal (no audible leak at an airway pressure of 20cm H2O), or to the maximum recommended volume. Cuff and airway pressures were measured continuously. A protocol was followed for repeated or failed attempts. 115 of the proposed 130 patients were r...
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Southern African Journal of Anaesthesia and Analgesia
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.